Additional Information
| Weight | 8.11 oz |
|---|---|
| Dimensions | 1.38 × 3.10 × 1.82 in |
| Features | โข Resolution of ยตrad |
| Applications | โข High-precision Geotech |
The Jewell LCF-300 Series flexure suspension servo fluid damped inclinometer is a ยฑ1ยฐ to ยฑ90ยฐ device designed for applications where high levels of shock and vibration are present. LCF-300 units are characterized by excellent turn on repeatability and very low hysteresis. It is available in three different versions (D, L, or S), varying input and output, and with either a 6 pin connector or a pin-terminal option in order to meet your specific needs.
| Weight | 8.11 oz |
|---|---|
| Dimensions | 1.38 × 3.10 × 1.82 in |
| Features | โข Resolution of ยตrad |
| Applications | โข High-precision Geotech |
Static/Dynamic
| Input Range (ยฐ) | ยฑ1.0 | ยฑ3.0 | ยฑ14.5 | ยฑ30.0 | ยฑ60.0 | ยฑ90.0 |
| Full Range Output (VDC, FRO ยฑ5%)1 | ยฑ5 | ยฑ5 | ยฑ5 | ยฑ5 | ยฑ5 | ยฑ5 |
| Non Linearity (%FRO max)2 | 0.05 | 0.02 | 0.02 | 0.02 | 0.02 | 0.05 |
| Scale Factor (Volts/g, nominal) | 285.5 | 95.5 | 20.0 | 10.0 | 5.8 | 5.0 |
| Scale Factor Temperature Sensitivity (PPM/ยฐC maximum) | 350 | 300 | 100 | 60 | 60 | 60 |
| Bandwidth (-3dB, Hz, nominal) | 0.5 | 2 | 15 | 20 | 30 | 30 |
| Transverse Axis Misalignment (ยฐ maximum) | ยฑ0.25 | ยฑ0.25 | ยฑ0.50 | ยฑ0.50 | ยฑ0.50 | ยฑ0.50 |
| 0ยฐ Output (Volts Range) | ยฑ0.10 | ยฑ0.04 | ยฑ0.02 | ยฑ0.02 | ยฑ0.02 | ยฑ0.02 |
| 0ยฐ Output Temperature Sensitivity (Volts/ยฐC maximum) | 0.015 | 0.005 | 0.001 | 0.0005 | 0.0004 | 0.0003 |
| Resolution & Threshold (ยต)3 | 1 | 1 | 1 | 1 | 1 | 1 |
Electrical
| Number of Axes | 1 |
| Input Voltage Range (VDC) | ยฑ12 to ยฑ18 |
| Noise (Vrms, maximum) | 0.002 |
| Input Current (mA, maximum) | ยฑ15 |
| Output Impedance (Ohms, nominal) | 10 |
Environmental & Enclosure
| Operating Temperature Range | -40ยฐC to +80ยฐC |
| Storage Temperature Range | -60ยฐC to +90ยฐC |
| Shock | 1500g, 0.5 msec, ยฝ sine |
| Seal | IP65 |
Notes:
1: Full Range is defined "from negative full input angle to positive full input angle."
2: Nonlinearity is specified as deviation of output referenced to theoretical sine function value, independent of misalignment.
3: Full Resolution is achieved with noise reduction techniques
Static/Dynamic
| Input Range (ยฐ) | ยฑ1.0 | ยฑ3.0 | ยฑ14.5 | ยฑ30.0 | ยฑ60.0 | ยฑ90.0 |
| Full Range Output (mA ยฑ1%)1 | 4-20 | 4-20 | 4-20 | 4-20 | 4-20 | 4-20 |
| Non-linearity (%FRO max)2 | 0.05 | 0.03 | 0.03 | 0.03 | 0.03 | 0.05 |
| Scale Factor (mA/g, nominal) | 458.4 | 152.9 | 32.0 | 16.0 | 9.2 | 8.0 |
| Scale Factor Temperature Sensitivity (PPM/ยฐC maximum) | 350 | 300 | 100 | 60 | 60 | 60 |
| Bandwidth (-3dB, Hz, nominal) | 0.5 | 2 | 15 | 20 | 30 | 30 |
| Transverse Axis Misalignment (ยฐ maximum) | ยฑ0.25 | ยฑ0.25 | ยฑ0.50 | ยฑ0.50 | ยฑ0.50 | ยฑ0.50 |
| 0ยฐ Output (mA) | 12 ยฑ0.6 | 12 ยฑ0.6 | 12 ยฑ0.3 | 12 ยฑ0.3 | 12 ยฑ0.3 | 12 ยฑ0.3 |
| 0ยฐ Output Temperature Sensitivity (Volts/ยฐC maximum) | 0.024 | 0.01 | 0.002 | 0.001 | 0.001 | 0.0008 |
| Resolution & Threshold (ยต)3 | 1 | 1 | 1 | 1 | 1 | 1 |
Electrical
| Number of Axes | 1 |
| Input Voltage Range (VDC) | 20 to 30 |
| Noise (Vrms, maximum) | 0.002 |
| Input Current (mA, maximum) | 40 |
Environmental & Enclosure
| Operating Temperature Range | -55ยฐC to +85ยฐC |
| Storage Temperature Range | -60ยฐC to +90ยฐC |
| Shock | 500g, 1 msec, ยฝ sine |
| Seal | IP65 |
Notes:
1: Full Range is defined "from negative full input angle to positive full input angle."
2: Nonlinearity is specified as deviation of output referenced to theoretical sine function value, independent of misalignment.
3: Full Resolution is achieved with noise reduction techniques
Static/Dynamic
| Input Range (ยฐ) | ยฑ1.0 | ยฑ3.0 | ยฑ14.5 | ยฑ30.0 | ยฑ60.0 | ยฑ90.0 |
| Full Range Output (VDC ยฑ0.5%)1 | 0-5 | 0-5 | 0-5 | 0-5 | 0-5 | 0-5 |
| Non-linearity (%FRO max)2 | 0.05 | 0.04 | 0.02 | 0.02 | 0.02 | 0.02 |
| Scale Factor (Volts/g, nominal) | 143.2 | 47.8 | 10.0 | 5.0 | 2.9 | 2.5 |
| Scale Factor Temperature Sensitivity (PPM/ยฐC maximum) | 350 | 150 | 100 | 60 | 60 | 60 |
| Bandwidth (-3dB, Hz, nominal) | 0.5 | 2 | 15 | 20 | 30 | 30 |
| Transverse Axis Misalignment (ยฐ maximum) | 0.25 | 0.50 | 0.50 | 0.50 | 1.0 | 1.0 |
| 0ยฐ Output (mA) | 2.48 to 2.52 | 2.48 to 2.52 | 2.48 to 2.52 | 2.48 to 2.52 | 2.48 to 2.52 | 2.48 to 2.52 |
| 0ยฐ Output Temperature Sensitivity (Volts/ยฐC maximum) | 0.0015 | 0.005 | 0.001 | 0.0005 | 0.0004 | 0.0003 |
| Resolution & Threshold (ยต)3 | 1 | 1 | 1 | 1 | 1 | 1 |
Electrical
| Number of Axes | 1 |
| Input Voltage Range (VDC) | +9 to +18 |
| Noise (Vrms, maximum) | 0.002 |
| Input Current (mA, maximum) | 40 |
| Output Impedance (Ohms, nominal) | 1 |
Environmental & Enclosure
| Operating Temperature Range | -40ยฐC to +80ยฐC |
| Storage Temperature Range | -60ยฐC to +90ยฐC |
| Shock | 1500g, .05 msec, ยฝ sine |
| Seal | IP65 |
Notes:
1: Full Range is defined "from negative full input angle to positive full input angle."
2: Nonlinearity is specified as deviation of output referenced to theoretical sine function value, independent of misalignment.
3: Full Resolution is achieved with noise reduction techniques